Reflectance pulse oximetry is a new method to monitor the fetus during
labour in order to get a better assessment of the fetus's condition t
han by CTG interpretation alone. In this study the method is investiga
ted with regard to its prognostic value concerning fetal outcome and c
omplications possibly due to its application: Increased uterine bleedi
ng, umbilical cord complications, amniotic infection syndrome. Arteria
l oxygen saturation was continuously recorded during labour in a total
of 122 patients with a gestational age between 37 and 42 weeks and co
rrelated to the umbilical cord-pH post partum and, if available, to in
tra partum fetal blood gas analysis. 30% seems to be a threshold value
below which oxygen saturation should not drop for a longer period of
time. A post partum pH < 7,2 can be expected in more than 50% of the c
ases in which the saturation remains below this value for more than te
n minutes. The application of the fetal pulse oximeter sensor did not
affect maternal or fetal morbidity.