DESIGN PARAMETERS OF DUAL-STAGE ION REFLECTRONS

Citation
Ti. Wang et al., DESIGN PARAMETERS OF DUAL-STAGE ION REFLECTRONS, Review of scientific instruments, 65(5), 1994, pp. 1585-1589
Citations number
10
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
65
Issue
5
Year of publication
1994
Pages
1585 - 1589
Database
ISI
SICI code
0034-6748(1994)65:5<1585:DPODIR>2.0.ZU;2-#
Abstract
Exact solutions of the second-order differential equation describing t he energy dependence of ion flight time in a reflectron time-of-flight mass spectrometer are derived, from which numerical evaluation of the construction and operation parameters of a dual-stage reflectron can easily be performed. Considerations in choosing the design parameters are discussed.