A rapid, accurate, precise and fully automated flow-injection method i
s described for the determination of 0.02-0.50 mu g ml(-1) cortisone,
0.02-1.00 mu g ml(-1) hydrocortisone and 0.20-5.00 mu g ml(-1) dexamet
hasone in aqueous solutions and of 1.0-20.0 mu g ml(-1) prednisolone,
methylprednisolone, progesterone, corticosterone and testosterone and
2.0-20.0 mu g ml(-1) betamethasone in 20% (v/v) acetonitrile solutions
. The method is based on the sensitizing effect of each steroid on the
chemiluminogenic cerium(IV)-sulphite reaction. Limits of detection ar
e in the ranges 0.013-0.019 and 0.05-4.00 mu g ml(-1) for aqueous and
acetonitrile solutions, respectively. Common excipients do not interfe
re greatly and the mean recovery of steroids from commercial formulati
ons was 102.6% (range 94.2-109.0%). The method allows the measurement
of up to 100 samples per hour and can be used for content uniformity t
ests and the results were in good agreement with the official method f
or commercial formulations with a relative error < 4%.