J. Hlavay et al., FIBEROPTIC BIOSENSOR FOR HYPOXANTHINE AND XANTHINE BASED ON A CHEMILUMINESCENCE REACTION, Biosensors & bioelectronics, 9(3), 1994, pp. 189-195
Fibre-optic biosensors were constructed for determination of hypoxanth
ine and xanthine. Xanthine oxidase and peroxidase were immobilized on
different preactivated membranes which were subsequently mounted onto
the tip of a fibre-optic bundle. The H2O2 generated by the reaction of
hypoxanthine and xanthine oxidase was measured by chemiluminescence (
CL) detection using luminol and peroxidase. A linear calibration curve
of the sensors in the range of 1-316 muM hypoxanthine and 3.1-316 muM
xanthine, respectively, with a detection limit of 0.55 muM hypoxanthi
ne was obtained. Recovery of hypoxanthine ranged between 91 and 102%.