TIGHTENING CHARACTERISTICS FOR SCREWED JOINTS IN OSSEOINTEGRATED DENTAL IMPLANTS

Citation
Rl. Burguete et al., TIGHTENING CHARACTERISTICS FOR SCREWED JOINTS IN OSSEOINTEGRATED DENTAL IMPLANTS, The Journal of prosthetic dentistry, 71(6), 1994, pp. 592-599
Citations number
11
Categorie Soggetti
Dentistry,Oral Surgery & Medicine
ISSN journal
00223913
Volume
71
Issue
6
Year of publication
1994
Pages
592 - 599
Database
ISI
SICI code
0022-3913(1994)71:6<592:TCFSJI>2.0.ZU;2-I
Abstract
The significance of tightening abutment screws and gold cylinders to o sseointegrated fixtures with the correct torque is demonstrated, and a simple relationship between applied torque and screw preload is deriv ed by use of mechanical engineering principles. The principles of a nu mber of tightening methods are outlined and assessments made of their accuracy. The difference between optimum and design torque is highligh ted. The necessity and means of achieving optimum torque to ensure a r eliable joint in clinical practice is discussed.