Rp. Hugtenburg et al., TOTAL-BODY IRRADIATION ON AN ISOCENTRIC LINEAR-ACCELERATOR - A RADIATION OUTPUT COMPENSATION TECHNIQUE, Physics in medicine and biology, 39(5), 1994, pp. 783-793
A treatment technique for total-body irradiation (TBI) is proposed tha
t combines arc therapy with dynamic output control to achieve high-gra
de dose uniformity. The patient lies on a low couch and receives expos
ure in the prone and supine positions from a modulated arcing beam. Th
e technique has been validated using a personal computer to control th
e linear accelerator and we demonstrate that only minor alterations to
current dynamic therapy systems would be required. We have examined t
he practical application of this treatment with emphasis on methods of
conformal therapy where an optimized dose distribution is prepared fr
om a matrix of caliper measurements taken from the patient. This techn
ique provides a means for regular TBI treatment on a computer-controll
ed linear accelerator that is easy to set up, requires short exposure
times and is comfortable for the patient.