INVESTIGATION OF THE IMAGING PROPERTIES OF AN X-RAY-FILM SCANNER

Citation
W. Doler et al., INVESTIGATION OF THE IMAGING PROPERTIES OF AN X-RAY-FILM SCANNER, Physics in medicine and biology, 39(5), 1994, pp. 917-922
Citations number
9
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
39
Issue
5
Year of publication
1994
Pages
917 - 922
Database
ISI
SICI code
0031-9155(1994)39:5<917:IOTIPO>2.0.ZU;2-K
Abstract
In this note, the imaging properties of a CCD x-ray film scanner were investigated. The relationship between optical density and the logarit hm of output pixel value provides information on die linearity of the system. The resolution properties of the scanner can be characterized by the presampling modulation transfer function which was calculated f rom a set of line spread functions with various alignments relative to die sampling grid. Our results show that the scanner is linear up to 2.0 optical density units, which leads to restrictions in digitizing x -ray films with higher optical density. The measured MTF shows the goo d spatial resolution of the CCD scanner which is demonstrated on an ex ample.