CALIBRATION POINT FOR ELECTRON IONIZATION MS MS SPECTRA MEASURED WITHMULTIQUADRUPOLE MASS SPECTROMETERS

Citation
Kr. Mohan et al., CALIBRATION POINT FOR ELECTRON IONIZATION MS MS SPECTRA MEASURED WITHMULTIQUADRUPOLE MASS SPECTROMETERS, Journal of the American Society for Mass Spectrometry, 5(6), 1994, pp. 576-582
Citations number
19
Categorie Soggetti
Chemistry Physical","Chemistry Analytical",Spectroscopy
ISSN journal
10440305
Volume
5
Issue
6
Year of publication
1994
Pages
576 - 582
Database
ISI
SICI code
1044-0305(1994)5:6<576:CPFEIM>2.0.ZU;2-6
Abstract
A protocol for establishing standard instrument conditions for measure ment of product ion MS/MS spectra from parent ions produced by electro n ionization is presented. Within this protocol, the ion at m/z 231 (C 5F9+) from perfluorokerosene or perfluorotributylamine is selected as the parent ion and subjected to collision-induced dissociation. The re lative intensities of product ions at m/z 69, 131, and 181 are monitor ed as a function of collision energy while keeping the target gas pres sure constant within the range of 10(-4)-10(-6) torr (measured), or a beam attenuation of approximately 30-70%. The collision energy at whic h the ion intensities for product ions at m/z 69 and 181 are equal is defined as the calibration point at that collision gas pressure; the i ntensity of the ion at m/z 131 is very close to this value as well. El ectron ionization MS/MS spectra taken at the calibration point using t wo different multiquadrupole instruments show good reproducibility for several test compounds. The high degree of similarity may aid in the establishment of a MS/MS spectral library.