U. Stumper, INTERNATIONAL COMPARISON GT-RF 75-A-11 - RF POWER AT 12 GHZ, 14 GHZ AND 17 GHZ IN 50-OMEGA COAXIAL LINE, IEEE transactions on instrumentation and measurement, 43(1), 1994, pp. 3-6
This paper deals with the results of an international comparison measu
rement of the effective efficiency E of commercial 50 OMEGA/7 mm coaxi
al thermistor RF power standards fitted with PC-7 precision connectors
. The measurement frequencies were 12 GHz, 14 GHz and 17 GHz. The nati
onal laboratories of metrology of eight countries (Australia, Canada,
France, Hungary, Italy, the Netherlands, USA and Germany) participated
in this comparison which was organized by the Radio Frequency Working
Group of the Comite Consultatif d' Electricite (CCE). The Physikalisc
h-Technische Bundesanstalt (PTB, Germany) acted as the pilot laborator
y. The standards showed only a small linear negative drift in E, as fo
und by linear regression analysis. For most of the mean values of E re
ported by the laboratories, the deviations from the regression values
were small (not larger than 0.004); only in two cases were the deviati
ons 0.006 and 0.0085.