A SUPERCONDUCTING CYCLOTRON FOR NEUTRON RADIATION-THERAPY

Citation
Rl. Maughan et al., A SUPERCONDUCTING CYCLOTRON FOR NEUTRON RADIATION-THERAPY, Medical physics, 21(6), 1994, pp. 779-785
Citations number
19
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
21
Issue
6
Year of publication
1994
Pages
779 - 785
Database
ISI
SICI code
0094-2405(1994)21:6<779:ASCFNR>2.0.ZU;2-8
Abstract
The physical and clinical specifications of a neutron therapy facility utilizing a superconducting cyclotron are presented. The cyclotron an d its support system are described. Details of the operation of the cy clotron in a hospital environment are given; the requirements of the h elium liquifier and cryogenic system are described together with a sum mary of its mode of operation. The simplicity of the cyclotron control system is discussed. The physical characteristics of the neutron beam are described. The central axis percent depth dose curve is equivalen t to that of a 4 MV x-ray beam. The depth of maximum dose occurs at ap proximately 9 mm depth and the surface dose is between 40% and 45%. Th e multirod collimator allows for the production of irregularly shaped fields of size up to 26.5 x 30 cm, without excessive exposure to opera ting personnel.