O. Kaltenbrunner et al., PREDICTION OF THE PREPARATIVE CHROMATOGRAPHY PERFORMANCE WITH A VERY SMALL COLUMN, Journal of chromatography, 760(1), 1997, pp. 41-53
Citations number
18
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Practical considerations lead to the use of small columns in preparati
ve chromatography. In these applications extra column effects may over
whelm all broadening effects. The influence of the extra column effect
s on the band broadening and migration is demonstrated by simulation.
The model assumes linearity of extra column effects. Consequently, the
se effects could be described by the convolution of the initial applie
d band width, an exponential decay and Gaussian broadening. An attempt
has been made to discriminate between the column, pre- and postcolumn
broadening.