DESIGN CONSIDERATIONS IN ENERGY-RESOLVED TIME-OF-FLIGHT MASS-SPECTROMETRY

Citation
Ae. Giannakopulos et al., DESIGN CONSIDERATIONS IN ENERGY-RESOLVED TIME-OF-FLIGHT MASS-SPECTROMETRY, International journal of mass spectrometry and ion processes, 131, 1994, pp. 67-86
Citations number
64
Categorie Soggetti
Spectroscopy,"Physics, Atomic, Molecular & Chemical
ISSN journal
01681176
Volume
131
Year of publication
1994
Pages
67 - 86
Database
ISI
SICI code
0168-1176(1994)131:<67:DCIETM>2.0.ZU;2-M
Abstract
We report the design, construction and performance of an energy resolv ed time-of-flight (ER-TOF) mass spectrometer. The system consists of a linear TOF coupled in parallel to the electric sector of a large doub le-focussing magnetic-sector mass spectrometer. The use of lensing bet ween the TOF and the electric sector is discussed with respect to the required beam conditions. Ion transmission through the electric sector is treated theoretically, especially with respect to time-of-flight. Preliminary experimental results from the instrument using matrix-assi sted laser desorption-ionisation are presented and discussed, with par ticular regard to the fragmentation of large ions.