CHARACTERIZATION, STANDARDIZATION AND EXPERIENCES WITH KONE ISE FOR MG2+

Citation
A. Lewenstam et al., CHARACTERIZATION, STANDARDIZATION AND EXPERIENCES WITH KONE ISE FOR MG2+, Scandinavian journal of clinical & laboratory investigation, 54, 1994, pp. 37-43
Citations number
6
Categorie Soggetti
Medicine, Research & Experimental
ISSN journal
00365513
Volume
54
Year of publication
1994
Supplement
217
Pages
37 - 43
Database
ISI
SICI code
0036-5513(1994)54:<37:CSAEWK>2.0.ZU;2-P
Abstract
The principles of a fully automated potentiometric determination of io nized magnesium in human blood serum are presented. By incorporating a magnesium selective electrode, into a commercially available six chan nel electrolyte analyzer (Microlyte 6, KONE Instruments, Finland), det ermination of the biologically-active, ionized fraction of magnesium h as been made possible. In this paper, the analytical performance of th e measurement system is presented and determination of ionized magnesi um in real patient samples is discussed with special attention to the significance of this new parameter.