ANALYSIS OF SERUM ELEMENTS AND THE CONTAMINATIONS FROM DEVICES USED FOR SERUM PREPARATION BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY
Citation
H. Morita et al., ANALYSIS OF SERUM ELEMENTS AND THE CONTAMINATIONS FROM DEVICES USED FOR SERUM PREPARATION BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY, Science of the total environment, 151(1), 1994, pp. 9-17
Categorie Soggetti
Environmental Sciences
SICI code
0048-9697(1994)151:1<9:AOSEAT>2.0.ZU;2-4
Abstract
Two different samples of serum were prepared from a blood specimen by
using two types of serum-separation tubes. Each serum sample was dilut
ed by 100 times with deionized and sub-boiling distilled water. Thirty
-six elements in the serum and exudate, from devices used for the seru
m preparation, were determined by inductively coupled plasma mass spec
trometry (ICP-MS). From the comparison of elemental concentrations in
two groups of 11 serum samples, it was found that the exudate from ser
um-separation tubes, as well as from disposable stainless steel needle
s, had serious effects on the elemental concentrations in the serum. M
eans for the concentrations of Li, Co, Ga, Cd, Sn, Sb, Ba, La, and Ce
in the two serum groups were statistically different from each other.
The concentrations of Cr, Mn, Fe, and Mo in both groups were apparentl
y higher than those reported in the literature, suggesting contaminati
on from the disposable stainless steel needle and spectral interferenc
es due to molecular species produced in the argon plasma.