Abstract
We have investigated the influence of sample size on the calibration o f a vibrating sample magnetometer. As expected, when measurements and calibration are carried out on samples with the same shape the results agree within the experimental uncertainty. For two cylindrical sample s of nickel, one with length and diameter 2.0 and 2.0 mm, and another with 5.3 and 1.0 mm respectively, the magnetometer signal varies with the length, in good agreement with numerical calculations.