RESPONSE OF TAPS TO MONOCHROMATIC PHOTONS WITH ENERGIES BETWEEN 45 AND 790 MEV

Citation
Ar. Gabler et al., RESPONSE OF TAPS TO MONOCHROMATIC PHOTONS WITH ENERGIES BETWEEN 45 AND 790 MEV, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 346(1-2), 1994, pp. 168-176
Citations number
21
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
346
Issue
1-2
Year of publication
1994
Pages
168 - 176
Database
ISI
SICI code
0168-9002(1994)346:1-2<168:ROTTMP>2.0.ZU;2-T
Abstract
The Two Arm Photon Spectrometer TAPS - comprising 384 plastic-BaF2 sci ntillator telescopes - was tested with monochromatic photons in the en ergy range between 45 and 790 MeV. The energy resolution for a collima ted photon beam hitting the central detector module was determined to sigma/E = 0.59% X E(gamma)-1/2 + 1.9% (E, given in GeV). For the the f ast scintillation component alone sigma/E = 0.79% X E(gamma)-1/2 + 1.8 % has been measured. The position resolution of the point of impact am ounts to DELTAx = 2 cm (FWHM) at the highest energies which correspond s to 30% of the diameter of an individual module. Monte Carlo simulati ons using the code GEANT3 are in good agreement with the experimental results.