L. Skillern et al., A COMPARATIVE-STUDY OF THE FETAL ELECTROCARDIOGRAM RECORDED BY THE STAN AND NOTTINGHAM SYSTEMS, British journal of obstetrics and gynaecology, 101(7), 1994, pp. 582-586
Objective To compare the T:QRS ratio record by the STAN and Nottingham
fetal electrocardiogram (FECG) monitors. Design Prospective observati
onal study. Setting London teaching hospital delivery suite and resear
ch unit. Methods and subjects The T:QRS ratios generated by the STAN a
nd Nottingham FECG monitors were simultaneously recorded and compared
using signals generated from a computer-produced ECG signal and signal
s from 11 term fetuses recorded during labour. Results There was an ac
ceptable level of agreement between the two systems with the computer-
generated signals, but it was not clinically acceptable with the signa
ls from the fetuses recorded during labour. Disagreements in the T:QRS
values were probably due to differences in the reference points for t
he measurement of the S-T segment and T-wave height. Conclusion The di
fferent points of reference for measurement of S-T segment and T-wave
height can explain poor agreement between the two methods of FECG wave
form analysis. The suggested adopted points of reference are those cor
responding to adult electrocardiographic methodology.