CATASTROPHIC FAILURE OF THE POLYETHYLENE LINER OF UNCEMENTED ACETABULAR COMPONENTS

Citation
Dj. Berry et al., CATASTROPHIC FAILURE OF THE POLYETHYLENE LINER OF UNCEMENTED ACETABULAR COMPONENTS, Journal of bone and joint surgery. British volume, 76B(4), 1994, pp. 575-578
Citations number
13
Categorie Soggetti
Surgery,Orthopedics
ISSN journal
0301620X
Volume
76B
Issue
4
Year of publication
1994
Pages
575 - 578
Database
ISI
SICI code
0301-620X(1994)76B:4<575:CFOTPL>2.0.ZU;2-Q
Abstract
Ten cases are described of catastrophic failure of the polyethylene li ner of three different designs of uncemented acetabular component. Fai lure occurred as a result of either 'wearthrough' to the metal backing , liner fracture or a combination of both, at a mean of 4.6 years afte r implantation (2 to 7.6). At revision there was metallosis in all hip s and osteolysis of the femur or the pelvis in six. Catastrophic failu re was seen only in cups with a minimum polyethylene thickness of less than 5 mm.