A NEW METHOD OF REAL-TIME ENDOSCOPIC MEASUREMENT WITH AN ELECTRIC CATHETER
Citation
T. Wakabayashi et al., A NEW METHOD OF REAL-TIME ENDOSCOPIC MEASUREMENT WITH AN ELECTRIC CATHETER, Endoscopy, 26(5), 1994, pp. 466-469
Categorie Soggetti
Gastroenterology & Hepatology
SICI code
0013-726X(1994)26:5<466:ANMORE>2.0.ZU;2-F
Abstract
A new method for measuring distances using electrically charged jets o
f water is described, and its precision examined. When combined with e
ndoscopy, this method can be utilized to measure the lengths and areas
of objects inside human body cavities. The apparatus used consists of
a catheter with three metal needles at the tip, a personal computer,
and a commercially available videoendoscopy system. To test the system
, a regulated voltage was passed between the catheter tip and a wooden
board, and two salt solutions of different concentrations were ejecte
d at three different angles from the catheter to the board. The coordi
nates of the board were then computed from the amperages of the curren
ts passing through the three jets of saline, and optical distortion wa
s corrected using the computer. The length and area of an object could
be measured rapidly by setting electrically the coordinates to match
the endoscopic image. Lengths ranging from 10 to 50 mm were measured w
hile varying the distance between catheter tip and board (10, 20, and
30 mm), as well as the angle formed by the catheter and the board (45,
60, and 90 degrees). The mean measurement error was less than 5.6 % a
nd accuracy was relatively independent of the angle between catheter a
nd target.