RAPID SCREENING TECHNIQUE FOR TETRACHLORODIBENZO-P-DIOXINS IN COMPLEXENVIRONMENTAL MATRICES BY GAS-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRYWITH AN ION-TRAP DETECTOR
Jb. Plomley et al., RAPID SCREENING TECHNIQUE FOR TETRACHLORODIBENZO-P-DIOXINS IN COMPLEXENVIRONMENTAL MATRICES BY GAS-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRYWITH AN ION-TRAP DETECTOR, Organic mass spectrometry, 29(7), 1994, pp. 372-381
A capillary column gas chromatographic/ion-trap tandem mass spectromet
ric (GC/MS/MS) technique was applied to the determination of tetrachlo
rodibenzo-p-dioxins (TCDDs) in complex environmental matrices. A detec
tion limit of 500 fg mul-1 injected, with a signal-to-noise ratio of 6
:1, was obtained when MS/MS was combined with broadband ion ejection o
f matrix ions and selected ion isolation. Calibration graphs were cons
tructed from extracted fragment ion mass chromatograms and were found
to be linear over the range 0.5-200 pg. Concentrations of total TCDDs
and of 2,3,7,8-TCDD were determined for extracts from fish, air and so
il; in most cases, the results found by GC/MS/MS were comparable to th
ose obtained by high-resolution MS and triple quadrupole MS.