THE DETERMINATION OF CATECHOLS IN THE PRESENCE OF ASCORBIC-ACID AND URIC-ACID BY FLOW-INJECTION ANALYSIS EMPLOYING A POTENTIOMETRIC DIBENZO-18-CROWN-6 ELECTRODE DETECTOR
Sk. Lunsford et al., THE DETERMINATION OF CATECHOLS IN THE PRESENCE OF ASCORBIC-ACID AND URIC-ACID BY FLOW-INJECTION ANALYSIS EMPLOYING A POTENTIOMETRIC DIBENZO-18-CROWN-6 ELECTRODE DETECTOR, Analytical letters, 27(11), 1994, pp. 2141-2151
A polished platinum disc was modified by electropolymerization of dibe
nzo-18-crown-6 (DB-18-C-6) onto the surface. This DB-18-C-6 electrode
has been used as a detector for catechol and catecholamines in a poten
tiometric-flow injection mode. The potentiometric-flow injection respo
nse of the DB-18-C-6 modified electrode exhibited a linear response fo
r catechol and catecholamine concentrations from 10(-2)-10(-6) M achie
ving detection limits as low as 0.5 x 10(-6) M. An interference study
was carried out with ascorbic acid and uric acid which are serious int
erferents using amperometric detectors. No significant interference wa
s observed for the presence of either of these two or mixture of these
two common interferents until they were in greater than a 10-fold exc
ess.