THE JOHNSON,ALLAN VOICE PROSTHESIS - A MODIFICATION OF THE BIVONA VOICE PROSTHESIS FOR IMMEDIATE POST-FITTING APHONIA AFTER SECONDARY TRACHEOESOPHAGEAL PUNCTURE

Citation
Pdr. Spraggs et al., THE JOHNSON,ALLAN VOICE PROSTHESIS - A MODIFICATION OF THE BIVONA VOICE PROSTHESIS FOR IMMEDIATE POST-FITTING APHONIA AFTER SECONDARY TRACHEOESOPHAGEAL PUNCTURE, Journal of Laryngology and Otology, 108(7), 1994, pp. 579-581
Citations number
5
Categorie Soggetti
Otorhinolaryngology
ISSN journal
00222151
Volume
108
Issue
7
Year of publication
1994
Pages
579 - 581
Database
ISI
SICI code
0022-2151(1994)108:7<579:TJVP-A>2.0.ZU;2-R
Abstract
The combination of a narrow oesophageal lumen and a hypertonic pharyng o-esophageal (P-E) segment following laryngectomy and secondary trache o-oesophageal puncture (TEP) can cause obstruction of a voice prosthes is and consequently prevent phonation. The 'Allan Johnson' modificatio n of the Bivona voice prosthesis incorporates a stainless steel slide and has been successfully used to remedy such a circumstance.