S. Faerestrand et al., LONG-TERM CLINICAL-PERFORMANCE OF A CENTRAL VENOUS OXYGEN-SATURATION SENSOR FOR RATE-ADAPTIVE CARDIAC PACING, PACE, 17(8), 1994, pp. 1355-1372
Rate adaptive ventricular pacemakers using central venous oxygen satur
ation (O2Sat) to control the pacing rate have been implanted in 14 pat
ients (mean age 71 years), with a mean follow-up period of 44 months (
range 2-63 months). In eight patients the pacemakers were replaced due
to signs of battery depletion after an implant duration of 39-58 mont
hs. During bicycle exercise testing the O2Sat decreased on average fro
m 61% +/- 4% at rest to 36% +/- 4% (P < 0.0001) at peak exercise, and
the maximum pacing rate was 122 +/- 5 beats/min. The time delay until
the 02Sat had dropped 10%, 65%, and 90% of the total reduction during
exercise was 4.8 +/- 0.9 seconds, 39.8 +/- 3.8 seconds, and 71.3 +/- 7
.5 seconds, respectively. The O2Sat decreased 9.4% +/- 2% (P < 0.005)
from resting supine to resting sitting. Oxygen breathing increased the
telemetered O2Sat from the pacemaker by 8.4 % +/-1% (P <0.001). Durin
g follow-up the O2Sats were relatively stable in 50% of the patients,
but demonstrated significant fluctuations in the others. At 1-year inv
asive follow-up O2Sat measured by the pacemaker decreased 22% +/- 2%,
and in blood samples from the right ventricle 22% +/- 2% from rest to
3 minutes exercise at 25 watts, There was a significant correlation be
tween O2Sat measured by the pacemaker and in blood samples from right
ventricle (n = 105; r = 0.73; P < 0.001). In two patients the O2Sat dr
opped significantly during pneumonia. In another patient episodes of a
ngina pectoris was associated with low O2Sat and a concomitant fast pa
cing rate.