SMALL SAMPLE DATING IN CHINA

Citation
Wj. Zhou et al., SMALL SAMPLE DATING IN CHINA, Radiocarbon, 36(1), 1994, pp. 47-49
Citations number
8
Categorie Soggetti
Geosciences, Interdisciplinary
Journal title
ISSN journal
00338222
Volume
36
Issue
1
Year of publication
1994
Pages
47 - 49
Database
ISI
SICI code
0033-8222(1994)36:1<47:SSDIC>2.0.ZU;2-I
Abstract
The Xi'an Laboratory of Loess and Quaternary Geology has developed a s mall sample C-14 dating facility consisting of a Wallac 1220 Quantulus (TM) liquid scintillation spectrometer, and a miniature benzene synthe sis line based on the synthesis procedures used at the Australian Nati onal University (ANU). This line can produce ca. 0.3-ml benzene sample s, which are then measured for C-14 activity using 0.3-ml Teflon vials developed by Wallac Oy. The counting performance of the Quantulus(TM) spectrometer using 0.3-ml vials has been evaluated, and a potential a ge limit of ca. 45,000 BP has been obtained for samples containing up to 250 mg carbon. This dating facility fills the gap between large sam ple (2.4-6 g carbon) and microsample (<1 mg carbon) handling to form a C-14 dating method sequence.