G. Naletto et al., RESPONSE ANALYSIS IN THE 300-ANGSTROM TO 2500-ANGSTROM SPECTRAL RANGEOF ULTRAVIOLET-ENHANCED CHARGE-COUPLED-DEVICES, Optical engineering, 33(8), 1994, pp. 2544-2552
The measurements of the ultraviolet response by some UV-enhanced CCDs
have been realized. In particular, the quantum efficiency of coronene-
and lumigen-coated and of back-illuminated ion-implanted CCDs have be
en measured in the 300 to 2500-angstrom spectral range. Very interesti
ng performances have been found, mainly for one ion-implanted CCD with
quantum efficiency values of more than 60% at 304 angstrom. Also some
measurements of the CCD uniformity response and spatial resolution ca
pability in this spectral region have been performed. The obtained res
ults are very promising for the possibility of using these semiconduct
or detectors in the ultraviolet spectral region also with very good pe
rformance.