A 2ND INTERCOMPARISON OF RN-222 MEASUREMENT SYSTEMS IN EUROPEAN LABORATORIES

Citation
Jcj. Dean et al., A 2ND INTERCOMPARISON OF RN-222 MEASUREMENT SYSTEMS IN EUROPEAN LABORATORIES, Applied radiation and isotopes, 47(9-10), 1996, pp. 835-840
Citations number
4
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging","Chemistry Inorganic & Nuclear
Journal title
Applied radiation and isotopes
ISSN journal
09698043 → ACNP
Volume
47
Issue
9-10
Year of publication
1996
Pages
835 - 840
Database
ISI
SICI code
0969-8043(1996)47:9-10<835:A2IORM>2.0.ZU;2-L
Abstract
A recent intercomparison exercise (established under the auspices of E UROMET) involving seventeen laboratories was conducted by NPL, who gen erated and standardised a set of glass-encapsulated Rn-222 samples and then dispatched a unique subset of the samples to each participant. T he reported measured activities of the samples were normalised to the NPL values. The spread of results, for the laboratories who had partic ipated in an earlier exercise, was of the order of +/-5%; the results of 12 of the participants agreed at the 1 sigma level. Copyright (C) 1 996 Elsevier Science Ltd.