B. Holmes et al., EVALUATION OF 2 BBL CRYSTAL SYSTEMS FOR IDENTIFICATION OF SOME CLINICALLY IMPORTANT GRAM-NEGATIVE BACTERIA, Journal of clinical microbiology, 32(9), 1994, pp. 2221-2224
Abstract
The BBL Crystal system (Becton Dickinson Microbiology Systems, Cockeys
ville, Md.) is a miniaturized bacterial identification method employin
g modified conventional and chromogenic substrates. Two products are c
urrently available, the Rapid Stool/Enteric ID Kit and the Enteric/Non
fermenter ID Kit, each comprising thirty tests. We report an evaluatio
n of both systems (using database version 1.1 for both) in the identif
ication of 51 gram-negative taxa likely to be encountered commonly in
the clinical laboratory. In all, 266 strains were tested in the Enteri
c/Nonfermenter ID Kit, and these represented 36 taxa of the family Ent
erobacteriaceae (188 strains), 5 oxidase-positive fermentative taxa (2
6 strains), and 10 nonfermentative taxa (52 strains). The majority of
these same strains (203 of 266) were also tested in the Rapid Stool/En
teric ID Kit. The Enteric/Nonfermenter ID Kit performed as follows: En
terobacteriaceae, 93% correct, 6% not identified, and 1% incorrect; ox
idase-positive fermenters, 88, 12, and 0%, respectively; and nonfermen
ters, 100% correct, although several only to the genus or group level.
The Rapid Stool/Enteric ID Kit gave the following results: Enterobact
eriaceae, 91% correct, 7% not identified, and 2% incorrect; oxidase-po
sitive fermenters, 80, 13, and 7%, respectively (but results were base
d on only 15 strains); and nonfermenters, 100% correct (but results we
re based on only 11 strains). We found the systems extremely easy and
rapid to use, and for the Enteric/Nonfermenter ID Kit an identificatio
n rate of 100% in 40 of 51 taxa was achieved, with corresponding figur
es of 29 of 39 taxa for the Rapid Stool/Enteric ID Kit.