SENSITIVE DETECTION OF IONIC ORGANOLEAD COMPOUNDS BY COUPLING HYDRIDEGENERATION (HG) WITH TRANSVERSELY HEATED GRAPHITE ATOMIZER ATOMIC-ABSORPTION SPECTROMETRY (THGA-AAS)

Citation
D. Erber et al., SENSITIVE DETECTION OF IONIC ORGANOLEAD COMPOUNDS BY COUPLING HYDRIDEGENERATION (HG) WITH TRANSVERSELY HEATED GRAPHITE ATOMIZER ATOMIC-ABSORPTION SPECTROMETRY (THGA-AAS), Fresenius' journal of analytical chemistry, 349(10-11), 1994, pp. 738-742
Citations number
13
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
349
Issue
10-11
Year of publication
1994
Pages
738 - 742
Database
ISI
SICI code
0937-0633(1994)349:10-11<738:SDOIOC>2.0.ZU;2-8
Abstract
The detection of ionic alkyllead compounds using the coupling of flow injection analysis system-hydride generation (FIAS-HG) with transverse ly heated graphite atomizer atomic absorption spectrometry (THGA-AAS) has been worked out. Very low limits of detection can be achieved if t he hydride products are enriched in the graphite furnace. Under optimi sed conditions (concentration of sodium borohydride, hydrogen peroxide and acidity as well as the furnace temperature) calibrations are carr ied out in the range of 0.1 to 5 mug/L. With a 1.5 mL sample loop, the limit of detection is calculated to be about 7 ng/L, but it can be lo wered to below 1 ng/L if larger sample volumes are used for the enrich ment.