A COMPARISON OF METHODS FOR PROMPT DETERMINATION OF POTENTIAL ALPHA-ENERGY CONCENTRATIONS

Citation
Ma. Duran et al., A COMPARISON OF METHODS FOR PROMPT DETERMINATION OF POTENTIAL ALPHA-ENERGY CONCENTRATIONS, Radiation protection dosimetry, 54(2), 1994, pp. 145-150
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
54
Issue
2
Year of publication
1994
Pages
145 - 150
Database
ISI
SICI code
0144-8420(1994)54:2<145:ACOMFP>2.0.ZU;2-V
Abstract
The object of this study was to verify experimentally an optimisation procedure used to reduce the total time needed to obtain an accurate e stimate of potential alpha energy concentration (PAEC) of Rn-222 proge ny for a grab sampling method with only one count of gross alpha activ ity. Samples of air were analysed for potential alpha energy at concen trations near 2.08 x 10(-6) J.m-3 (0.10 WL) using timing intervals of 7, 13, and 25 min. The measurements were compared graphically and stat istically to estimates of PAEC obtained using the modified Tsivoglou m ethod. No statistical differences in accuracy were found within or bet ween the three total timing intervals. Suggestions for practical field instrumentation are presented.