POTENTIOMETRIC SENSOR-BASED ON MODIFIED BLM ON RIGID SUPPORT

Citation
D. Xiao et al., POTENTIOMETRIC SENSOR-BASED ON MODIFIED BLM ON RIGID SUPPORT, Analytical letters, 27(12), 1994, pp. 2249-2259
Citations number
8
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032719
Volume
27
Issue
12
Year of publication
1994
Pages
2249 - 2259
Database
ISI
SICI code
0003-2719(1994)27:12<2249:PSOMBO>2.0.ZU;2-J
Abstract
In the present work, ammonium hexadecanoyl trioctyl iodide (AHTI), iod ine and N,N-(1,4-butylidene) thiuram disulfide (BITDS) modified BLM-s on Pt support have been constructed. The variation of potential during the process of BLM formation on a fresh cut tip surface have been obs erved and its potentiometric response toward iodide or copper ion have been investigated, The AHTI modified BLM on Pt support shows linear p otentiometric response for iodide ion in the range of 10(-6) - 10(-2) mol/L with a slope of 57 mV/pC. For the iodine modified BLM on Pt supp ort the linear response range for iodide ion is 10(-5)- 10(-2) mol/L w ith a slope of 55 mV/pC. The BITDS modified BLM on Pt support exhibits a Nernstian response for copper ion in the range of 10(-5)-10(-2) mol /L with a slope of 33 mV/pC.