OPTIMIZATION OF DETECTORS IN TIME-FOCUSING GEOMETRY FOR RTOF NEUTRON DIFFRACTOMETERS

Citation
Va. Kudryashev et al., OPTIMIZATION OF DETECTORS IN TIME-FOCUSING GEOMETRY FOR RTOF NEUTRON DIFFRACTOMETERS, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 93(3), 1994, pp. 355-361
Citations number
19
Categorie Soggetti
Physics, Nuclear","Nuclear Sciences & Tecnology","Instument & Instrumentation
ISSN journal
0168583X
Volume
93
Issue
3
Year of publication
1994
Pages
355 - 361
Database
ISI
SICI code
0168-583X(1994)93:3<355:OODITG>2.0.ZU;2-T
Abstract
The quality of a neutron time-of-flight diffractometer can be signific antly improved by setting up its detectors in time-focusing geometry. In order to fully exploit the time-focusing method, the nominal resolu tion of the spectrometer must be thoroughly investigated by analysing the contributions of several different factors. In this paper analytic formulae are given, which describe the different components contribut ing to the resolution function and the conditions which optimize the r esolution are derived. It is shown how the analytic integral equations can be solved numerically. The programs allow iterative computations to optimize a detector system composed of many individual detector ele ments, considering the specimen size and instrumental resolution. The flow diagram of the programs to calculate numerically the detector sys tem is briefly discussed and the precision of the individual optimal d etector position coordinates is given.