A HIGH-RESOLUTION VARIABLE-LINE-SPACING SPHERICAL-GRATING MONOCHROMATOR WITH A MOVABLE EXIT SLIT

Citation
R. Reininger et M. Bissen, A HIGH-RESOLUTION VARIABLE-LINE-SPACING SPHERICAL-GRATING MONOCHROMATOR WITH A MOVABLE EXIT SLIT, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 347(1-3), 1994, pp. 269-272
Citations number
13
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
347
Issue
1-3
Year of publication
1994
Pages
269 - 272
Database
ISI
SICI code
0168-9002(1994)347:1-3<269:AHVSM>2.0.ZU;2-0
Abstract
A design for a spherical-grating monochromator based on the rotation o f a variable-line-spacing grating and a movable exit slit in order to minimize the defocus, coma, and spherical aberrations is discussed. Th e resolution of the proposed monochromator has been investigated both analytically and by means of ray tracing. With a grating figure error of +/-0.1 arcsec, the resolving power ranges from 10(4) to 5 x 10(3) o ver the energy range 380-1260 eV. The total travel of the exit slit re quired for scanning this energy range is less than 250 mm.