Abstract
We have constructed the first working model of a novel diffraction geo metry for the high-resolution monochromatization of hard (5-30 keV) sy nchrotron radiation. The monochromator has an energy bandpass of 6.7 m eV FWHM and an angular acceptance of 22 murad (4.5 arcseconds) for 14. 4 keV X rays - an energy resolution DELTAE/E of 4.6 X 10(-7). With a s econd set of crystals, an energy bandpass of 23 meV FWHM and an angula r acceptance of 7 murad was achieved for 23.8 keV photons. The perform ance of the monochromator has been verified, in a nearly ideal measure ment, using nuclear-fluorescent X rays with an energy bandwidth on the order of 10 neV.