M. Bissen et al., EVALUATION OF AN OPTICAL BEAM-POSITION-MONITOR SYSTEM WITH CLOSED-LOOP STEERING CAPABILITIES, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 347(1-3), 1994, pp. 581-583
Imaging the synchrotron source profile onto the entrance slit of a mon
ochromator provides a stable and reproducible energy calibration which
is independent of the absolute position and drift of the electron bea
m. Potential electron-beam motions occurring during a fill result in a
loss of flux through the beamline. We have implemented two independen
t beam position monitors which can be used as sensors to steer the ver
tical entrance mirror in order to maintain a maximum flux through a sp
herical grating varied line-spacing monochromator beamline. The system
consists of a slotted plate photodiode which intercepts 2 mrad of syn
chrotron radiation next to the entrance mirror and a detector utilizin
g the photocurrents generated at the jaws of the entrance-slit assembl
y. Both monitors have a wide linear response range with a vertical pos
ition resolution of < 5 mum. The combination of two independent beam p
osition monitors allows an easy check on the mechanical and thermal st
ability of the entrance optical system as well as on the reproducibili
ty and long-term fluctuations of the electron-beam source during user
shifts. We will discuss the performance of the optical beam-position-m
onitor system and its implementation as a sensor in a closed-loop feed
back system to maintain maximum flux through the beamline.