PRACTICAL ASPECTS OF SHIELDED GRADIENT-COIL DESIGN FOR LOCALIZED IN-VIVO NMR-SPECTROSCOPY AND SMALL-SCALE IMAGING

Citation
Cd. Eccles et al., PRACTICAL ASPECTS OF SHIELDED GRADIENT-COIL DESIGN FOR LOCALIZED IN-VIVO NMR-SPECTROSCOPY AND SMALL-SCALE IMAGING, Magnetic resonance imaging, 12(4), 1994, pp. 621-630
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
0730725X
Volume
12
Issue
4
Year of publication
1994
Pages
621 - 630
Database
ISI
SICI code
0730-725X(1994)12:4<621:PAOSGD>2.0.ZU;2-E
Abstract
For a number of NMR applications the availability of screened gradient s is crucial to minimize the effect of eddy currents on the NMR signal . In this paper we review two types of shielded gradient design: the t arget field and minimum inductance methods. The two designs are compar ed and construction details are presented. A novel coil design constru cted with a double-sided primary and a single layered screen is outlin ed. Experimental results for both target field and minimum inductance coils are presented and compared.