Pmt. Pattynama et al., CARDIOVASCULAR MR-IMAGING - PRESSURE-GATING USING THE ARTERIAL-PRESSURE SIGNAL FROM A CONVENTIONAL FERROMAGNETIC MICROMANOMETER-TIP CATHETER, Magnetic resonance imaging, 12(3), 1994, pp. 531-534
We used the arterial pressure wave as a reliable trigger to synchroniz
e MR imaging to the cardiac cycle. Pressure readings and MR images rem
ained undistorted using a conventional ferromagnetic micromanometer-ti
p catheter. As a safe alternative to triggering to ventricular pressur
e, this approach allows varying the time relation between the trigger
and the cardiac cycle while maintaining the advantages of pressure-gat
ing.