A turbine flow sensor for MR flow experiments has been evaluated using
reference volume-flow rate measurements obtained using an electromagn
etic (EM) flow meter measurements and simultaneous phase contrast (PC)
MR acquisitions. After calibration, the device was found to have accu
racy (compared with the EM flow meter), linearity, and precision of be
tter than +/- 1%, +/- 3.5%, 3.5%, respectively, in constant flow mode
(0 to 30 ml s(-1)). The frequency response of the flow sensor was flat
(within +/- 10%) up to 13.9 Hz. Volume-flow rate measurements on cons
tant and simulated physiologic flow waveforms were in close agreement
with both the electromagnetic (EM) flow meter and the gated MR PC esti
mates.