REALIZATION OF THE ITS-90 AT IMGC IN THE RANGE OF LONG STEM PRTS

Citation
P. Marcarino et R. Dematteis, REALIZATION OF THE ITS-90 AT IMGC IN THE RANGE OF LONG STEM PRTS, Measurement techniques, 36(12), 1993, pp. 1406-1414
Citations number
13
Categorie Soggetti
Instument & Instrumentation",Engineering
Journal title
ISSN journal
05431972
Volume
36
Issue
12
Year of publication
1993
Pages
1406 - 1414
Database
ISI
SICI code
0543-1972(1993)36:12<1406:ROTIAI>2.0.ZU;2-6
Abstract
IMGC has established the fixed points required by the International Te mperature Scale of 1990 (ITS-90) for long-stem platinum resistance the rmometers (PRTs) at the highest accuracy level. The present paper desc ribes the fixed point apparatus and the measuring procedures used at I MGC for the accurate realization of each fixed point, and gives some d ata on the typical phase transitions thereby obtained. Moreover, since IMGC has developed both traditional and sealed cells for the realizat ion of the freesing or melting points of metals, the different types o f cells and the procedure used for their construction and preparation are also supplied. Finally, the reproducibilities of the freezing poin ts of Sn, Zn, Al, and Ag were evaluated by means of several determinat iaons with one standard thermometer.