OBSERVED REDOX INTERFERENCES OF POLY(PYRROLE)-BASED PERCHLORATE-SELECTIVE ELECTRODES

Citation
A. Hulanicki et al., OBSERVED REDOX INTERFERENCES OF POLY(PYRROLE)-BASED PERCHLORATE-SELECTIVE ELECTRODES, Electroanalysis, 6(7), 1994, pp. 604-605
Citations number
6
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
10400397
Volume
6
Issue
7
Year of publication
1994
Pages
604 - 605
Database
ISI
SICI code
1040-0397(1994)6:7<604:ORIOPP>2.0.ZU;2-M
Abstract
The responses of the poly(pyrrole)-based ion-selective perchlorate ele ctrode were investigated in both perchlorate and redox buffers media. Sensitivity of tested electrodes both to the changes of perchlorate io n concentrations and redox potential of the solution were observed. Sl opes were characteristic: -51.3 mV/dec in the case of perchlorate ion solutions and 61 mV/dec in the case of redox calibration. Potentiometr ic redox sensitivity of the tested poly(pyrrole) layers was observed i n presence of 1 M sodium perchlorate. Potentiometric redox sensitivity of the bare platinum electrode and perchlorate-sensitive electrode ba sed on poly(pyrrole) in presence of 0.1 M NaClO4 were compared.