A. Gaikwad et al., SENSITIVE DETERMINATION OF PERIODATE AND TARTARIC ACID BY STOPPED-FLOW CHEMILUMINESCENCE SPECTROMETRY, Analyst, 119(8), 1994, pp. 1819-1824
The stopped-flow technique was used for the determination of periodate
by;measuring the chemiluminescence produced on reaction with luminol,
enhanced by manganese(II) and triethanolamine. The method was also ap
plied to the determination of tartaric acid by oxidation with periodat
e. Under the optimum conditions,responses were linear between 1.0 x 10
(-6) and 2.5 x 10(-4) mol l(-1) periodate and from 5.0 x 10(-6) to 5.0
x 10(-5) mol l(-1) tartaric acid, with a relative standard deviation
below about 2% in both instances. The proposed method compares favoura
bly with existing alternatives in terms of sensitivity and sampling fr
equency. It was used for the determination of tartaric acid in pharmac
eutical preparations, the results being consistent with the manufactur
ers' certified contents.