Dm. Binkley, OPTIMIZATION OF SCINTILLATION DETECTOR TIMING SYSTEMS USING MONTE-CARLO ANALYSIS, IEEE transactions on nuclear science, 41(1), 1994, pp. 386-393
Monte Carlo analysis is used to model statistical noise associated wit
h scintillation-detector photoelectron emissions and photomultiplier t
ube operation. Additionally, the impulse response of a photomultiplier
tube, front-end amplifier, and constant-fraction discriminator (CFD)
is modeled so the effects of front-end bandwidth and constant-fraction
delay and fraction can be evaluated for timing-system optimizations.
Such timing-system analysis is useful for detectors having low photoel
ectron-emission rates, including Bismuth Germanate (BGO) scintillation
detectors used in Positron Emission Tomography (PET) systems. Monte C
arlo timing resolution for a BGO/photomultiplier scintillation detecto
r, CFD timing system is presented as a function of constant-fraction d
elay for 511-keV coincident gamma rays in the presence of Compton scat
ter. Monte Carlo results are in good agreement with measured results w
hen a tri-exponential BGO scintillation model is used. Monte Carlo sim
ulation is extended to include CFD energy-discrimination performance.
Monte Carlo energy-discrimination performance is experimentally verifi
ed along with timing performance (Monte Carlo timing resolution of 3.2
2 ns FWHM versus measured resolution of 3.30 ns FWHM) for a front-end
rise time of 10 ns (10-90%), CFD delay of 8 ns, and CFD fraction of 20
%.