Hw. Julius et al., TYPE TESTING OF PERSONAL DOSIMETERS FOR PHOTON ENERGY AND ANGULAR RESPONSE, Radiation protection dosimetry, 54(3-4), 1994, pp. 273-277
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
Performance criteria for personal dosemeters have been recently propos
ed in the CEC document 'Technical Recommendations for Monitoring Indiv
iduals Occupationally Exposed to External Radiation' (EUR 5287, Revise
d). The current experiment was carried out to demonstrate that TLD-bas
ed dosimetry systems can comply with these requirements. The emphasis
of the experiment was on the dosemeter response as a function of photo
n energy and angle of incidence of the radiation. The tests were based
on the ICRU operational quantities H(p) (0.07) and H(p) (10). All mea
surements were compared with the true dose in a 30 x 30 x 15 cm tissue
-equivalent slab phantom, calculated using appropriate conversion coef
ficients. For each combination of photon energy and angle, three dosem
eters of each system were irradiated simultaneously on a 30 x 30 x 15
cm PMMA phantom. No corrections were made for possible - but unknown!
- influence on the dosemeters from the difference in backscatter betwe
en tissue-equivalent material and PMMA.