A 2-ELEMENT TL DOSIMETER FOR THE MEASUREMENT OF H(P)(0.07) AND H(P)(10) IN MIXED PHOTON BETA-FIELDS

Citation
P. Ambrosi et al., A 2-ELEMENT TL DOSIMETER FOR THE MEASUREMENT OF H(P)(0.07) AND H(P)(10) IN MIXED PHOTON BETA-FIELDS, Radiation protection dosimetry, 54(3-4), 1994, pp. 295-298
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
54
Issue
3-4
Year of publication
1994
Pages
295 - 298
Database
ISI
SICI code
0144-8420(1994)54:3-4<295:A2TDFT>2.0.ZU;2-G
Abstract
A dosemeter badge with two TL detectors has been designed for the simu ltaneous measurement of H(p)(0.07) and H(p)(10). The performance of th e dosimetry system has been investigated for various X and beta ray fi elds with radiation incidence between 0-degree and 75-degrees. For arb itrary mixtures of the test radiation fields the response to H(p)(0.07 ) varies between 0.57 and 1.62. The response of the dosemeter to H(p)( 10) ranges between 0.39 and 1.60 for X rays. The thin TL detector has some stability problems which can probably be reduced by optimising th e annealing procedure.