LOW-NOISE SCINTILLATION DETECTORS WITH A P-47 THIN-LAYER SCREEN FOR ELECTRONS OF SEVERAL KEV

Citation
Z. Kajcsos et al., LOW-NOISE SCINTILLATION DETECTORS WITH A P-47 THIN-LAYER SCREEN FOR ELECTRONS OF SEVERAL KEV, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 93(4), 1994, pp. 505-515
Citations number
25
Categorie Soggetti
Physics, Nuclear","Nuclear Sciences & Tecnology","Instument & Instrumentation
ISSN journal
0168583X
Volume
93
Issue
4
Year of publication
1994
Pages
505 - 515
Database
ISI
SICI code
0168-583X(1994)93:4<505:LSDWAP>2.0.ZU;2-J
Abstract
The applicability of a low-noise scintillation detector (ScD) for the registration of electrons of several keV energy has been studied emplo ying photomultipliers (PM) of different types and sizes. With the appl ication of a sedimented P-47 scintillation screen, the values of the l ow-energy sensitivity limit and those of the light conversion coeffici ent were determined as about 2.7-4.7 keV and 2.8-6.6 photoelectrons/ke V, respectively, for the set of PM's (Philips-Valvo XP 2020, Philips-V alvo XP 2052, Philips-Valvo XP 2972, EMI 9124a) studied. It is conclud ed that such scintillation detectors might be used advantageously as e lectron counters in the range of E > 5 keV. Applications below this ki netic energy value are also feasible when applying a floating accelera tion of several kV to the ScD - a voltage much lower than the values r equired for Everhart-Thornley detectors.