DESIGN CONSIDERATIONS FOR PREAMPLIFIERS TO BE USED ON FAST MULTIWIRE GAS PROPORTIONAL COUNTER READOUT SYSTEMS

Citation
Ao. Jones et al., DESIGN CONSIDERATIONS FOR PREAMPLIFIERS TO BE USED ON FAST MULTIWIRE GAS PROPORTIONAL COUNTER READOUT SYSTEMS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 348(2-3), 1994, pp. 256-260
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
348
Issue
2-3
Year of publication
1994
Pages
256 - 260
Database
ISI
SICI code
0168-9002(1994)348:2-3<256:DCFPTB>2.0.ZU;2-T
Abstract
New high rate multiwire gas proportional counters for X-ray detection under development at Daresbury require low noise, fast preamplifiers t o achieve minimum dead time. We present a design for a fast preamplifi er with a rise time of 8 ns, noise better than 1300 electrons rms and a gain of 4.8 muV per electron. Results of a computer simulation run t o predict the optimum differentiate constant needed to ensure a short dead time are also presented. The application of this preamplifier on a fast multiwire linear detector is discussed and results are presente d.