Ra. Kroeger et al., CHARGE DIVISION READOUT OF A 2-DIMENSIONAL GERMANIUM STRIP DETECTOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 348(2-3), 1994, pp. 507-509
Germanium strip detectors combine superior spectroscopy typical of ger
manium detectors with good spatial resolution. Our work has been to de
velop a 2-dimensional position readout of a 5 x 5 strip X-Y detector.
The prototype device uses two capacitive charge division strings. Each
string is read out by two 13-bit ADCs, one on each end of the string.
The four data channels are stored as an event list for subsequent pro
cessing. We form a response map over the detector surface in order to
locate the position of each interaction with the spatial resolution of
the strip pitch, in our case 9 mm. Cross-talk and non-linearities are
removed using the response map. Energy resolution of 5-6 keV FWHM is
achieved between 60 and 662 keV across the full surface of the detecto
r.