OPERATION OF A BAF(2)-TMAE PROTOTYPE DETECTOR

Citation
K. Wells et al., OPERATION OF A BAF(2)-TMAE PROTOTYPE DETECTOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 348(2-3), 1994, pp. 600-606
Citations number
11
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
348
Issue
2-3
Year of publication
1994
Pages
600 - 606
Database
ISI
SICI code
0168-9002(1994)348:2-3<600:OOABPD>2.0.ZU;2-7
Abstract
A fully optimised BaF2-TMAE detector has been developed for use in Pos itron Emission Tomography (PET). As part of this work we have attempte d to optimise the performance of this technology with respect to effic iency, timing (resolution and deadtime), and spatial resolution for 51 1 keV photons. We report for the first time on the detailed operation of our detector. Under stable operating conditions over an 8 hour peri od, we consistently obtain an absolute detection efficiency of 23.5% f or an 8 mm thick crystal, simultaneously with 4 ns FWHM and 9.5 FWTM t iming resolution and a spatial resolution of 6 mm FWHM. We also discus s how these parameters affect the performance of a full-size camera ba sed on BaF2-TMAE detectors.