Ik. Perera et al., CONTINUOUS-WAVE LASER RESONANCE IONIZATION MASS-SPECTROMETRIC ANALYSIS OF STRONTIUM, International journal of mass spectrometry and ion processes, 132(3), 1994, pp. 225-238
Citations number
48
Categorie Soggetti
Spectroscopy,"Physics, Atomic, Molecular & Chemical
Resonance ionization mass spectrometric analysis of strontium has been
performed through a simple two colour 1 + 1 cw laser excitation schem
e, using a portion of the pump Ar+ ion laser to excite the non-resonan
t ionization step, and the first results of precise isotope ratio meas
urements of Sr are presented. These measurements made in an ion microp
robe, employing both thermal evaporation and primary ion beam sputteri
ng for producing neutral Sr atoms, have exhibited strong even-odd effe
cts for the Sr-87/Sr-86 ratio, with a near 10% depletion of the Sr-87
isotope, and the dependence of this anomalous isotopic effect on laser
output power has been studied. In sputter-induced RIMS experiments, p
rimary ion beams of oxygen, argon and xenon, were used for sputtering
pure Sr metal targets, and the calculated ionization yields are compar
ed with the results of a simple mathematical analysis of the resonance
ionization process. Possible reasons for the quantitative differences
observed between the measured and the predicted results are suggested
and discussed.