RUNAWAY PACEMAKER DURING HIGH-ENERGY NEUTRON RADIATION-THERAPY

Citation
Mh. Raitt et al., RUNAWAY PACEMAKER DURING HIGH-ENERGY NEUTRON RADIATION-THERAPY, Chest, 106(3), 1994, pp. 955-957
Citations number
5
Categorie Soggetti
Respiratory System
Journal title
ChestACNP
ISSN journal
00123692
Volume
106
Issue
3
Year of publication
1994
Pages
955 - 957
Database
ISI
SICI code
0012-3692(1994)106:3<955:RPDHNR>2.0.ZU;2-7
Abstract
Runaway pacemaker occurred in a patient undergoing high-energy neutron radiation therapy despite adherence to published safety guidelines. T he very low estimated dose of 0.9 Gy received by the pacemaker demonst rates the extreme sensitivity of integrated circuits to this new modal ity of radiation therapy.