Cg. Demaria et al., REACTIVATION OF AN IMMOBILIZED ENZYME REACTOR FOR THE DETERMINATION OF ACETYLCHOLINESTERASE INHIBITORS FLOW-INJECTION DETERMINATION OF PARAOXON, Analytica chimica acta, 295(3), 1994, pp. 287-296
A stopped-flow/flow-injection procedure is proposed for the spectropho
tometric determination of paraoxon based on the inhibition of immobili
zed acetylcholinesterase. The irreversibly inhibited enzyme is reactiv
ated with 1,1'-trimethylene-bis(4-formylpyridinium bromide) dioxime. T
he proposed procedure has a detection limit of 4 x 10(-9) M, a linear
calibration range of 9 x 10(-9)-5 x 10(-7) M, a R.S.D. of 1.6% (n = 10
) at 5.0 X 10(-8) M and a sample throughput frequency of 10 h(-1). The
enzyme reactor can be used for more than one thousand inhibition-reac
tivation cycles.